Chapter
7
Device check
WARNING
WARNING
WARNING
Selftest
118
Run the selftest and the system test in all of the
following situations:
▪
▪
▪
▪
▪
Make sure that you have carried out all work properly
in accordance with (→ "Preparations" p. 68).
Device malfunction!
Death or permanent injury of the patient
▪
A device check must be run once a day.
Proper state of the device is not monitored, system test and self test not
executed/skipped!
Death or permanent injury of the patient
▪
Run the self test and the system test:
Alarms at system start-up: Device malfunction!
Death or permanent injury of the patient
▪
Make sure no alarms are triggered at start-up.
This test is executed automatically when the device
is turned on.
(→ "Turn on" p. 120)
Make sure the environment is quiet.
The selftest (starts when the device is turned on) and
the associated hardware check must be run once a
day.
Once a day
Before the first start-up
After every servicing and/or repair
After the device has been moved to a different
location
After work on the central gas supply