High Level 2-Channel Lab Scope with built in Multimeter
Channel 1 marker
Shows the reference points of the displayed waveforms.
25
If there is no marker, the channel is not displayed.

3.1.2 Automatic Calibration

The automatic calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run the routine if the
ambient temperature changes by 10°C or more.
For accurate calibration, power on the MAX2CSCOPE and wait twenty minutes to ensure it is warmed
up. Calibration mainly includes simulation channel calibration, trigger voltage of trigger circuit calibration
and non-linear calibration of horizontal base line deviation.
Note : During auto calibration, please make sure no signal in input terminal of CH1 / CH2, otherwise it
may damage device.
Tap Menu -> User Settings, then tap Automatic Calibration.
Tap OK on the pop-up message box to start calibration. The LED on the MAX2CSCOPE starts flashing
and the prompt message "Calibrating, please wait" will appear on the upper left corner of the screen.
Note: While calibrating, no operations are allowed to perform until the calibration process is
successfully finished.
The self-calibration routine takes several minutes. After calibration is complete, the prompt message
"Correction successful" will pop up on the screen.
Tap OK to complete automatic calibration.

3.1.3 Connection

Testing Sensor:
1. Connect the BNC connector of the BNC to 4mm test lead to the CH1/CH2 (select number of channel
and channel No. according to actual situations), and plug the black (GND) and red (SIGNAL) 4mm
Same with "Channel 2
marker".