USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
Table 9 describes fault criteria for the Analyzer Alarms.
'A' refers to Laser 1 and 'B' refers to Laser 2. Not all analyzers are equipped
with 2 lasers.
Status
Sensor Read
Data Health
4
(A/B)
5
Pressure
6
HD Space
Mirror Health
7
(A/B)
Linelock
8
(A/B)
Signal Power
9
(A/B)
10
Maintenance
11
Data Health
12
(A/B)
13
Pressure
14
HD Space
Mirror Health
15
(A/B)
Linelock
16
(A/B)
Signal Power
17
(A/B)
18
Maintenance
19
20
Performance
Alarm Status
If the
section on page 65. If issue continues, please contact
icos.support@ca.abb.com
Table 9: Fault Criteria for Analyzer Alarms
Fault Condition
The laser fitting condition is poor. Occurs when
Fit is not optimal
fit is no longer working, peaks have been lost, or
spectrum is unknown.
Not within
Occurs when pressure is outside of the operating
operating range
range.
Limited hard drive
Occurs when the internal hard drive has < 10% of
space
space left. Delete unnecessary data files.
Mirrors have
Occurs when the ringdown time has degraded by
declined in
> 20% of the factory value. Mirror cleaning is
reflectivity
required.
Peak is outside
Occurs when linelock control voltage is no longer
control range
able to control.
Signal power has
Occurs when laser signal power has degraded by
degraded
> 20% of the factory value.
Maintenance
Occurs when the analyzer requires maintenance
needed now
(every 381 days).
CROSS OVER
The laser fitting condition is not optimal. Occurs
Fit is not optimal
when residuals of fit go above normal operational
values.
Occurs when the specified operational pressure is
Noisy
not optimal.
Occurs when the internal hard drive has < 20%
Low space
space left. Delete unnecessary data files.
Mirrors have
Occurs when the ringdown time has degraded by
declined in
> 10% of the factory value.
reflectivity
Occurs when linelock control voltage is
Peak is drifting
approaching control range limit.
Signal power is
Occurs when laser signal power has degraded by
degrading
> 10% of the factory value.
Maintenance
Analyzer maintenance will be needed soon (every
needed soon
360 days).
Dead Band
No warning/alarm
No warning/alarm
is Yellow or
Red,
Description
Maintenance
please refer to the