USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
Changing the Septa on the Syringe Injection Port
The septum on the syringe injection port requires periodic replacement. Depending on use,
a septum should last a minimum of 100 injections.
To replace the septum:
1.
Click the Rate button on the
2. The Operating Mode pop-up menu displays. (Figure 111)
3. Set the analyzer to
buttons under the
Figure 111: Rate Control Screen set to Slow Flow Mode
4. Unscrew the septum nut from the injection port as shown in Figure 112.
5. Remove the red septum with white Teflon coating from the inside of the septum
nut, using the provided septum puller. Discard the used septum.
6. Obtain a new septum from the provided package.
User Interface Control Bar
Slow
(Continuous Flow) mode by selecting one of the radio
Slow
option. (Figure 111)
Figure 112: Septum Nut with used septum
. (Figure 107)