USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
Batch Mode Processing
batch processing mode,
In
1.
Initiates the batch injection procedure.
2. Evacuates the cavity.
3. Flushes the cavity with zero air twice before requesting the sample.
4. Prompts you to inject sample (of >60 ml) gas into the syringe port.
a. You have 120 seconds to inject the sample.
If you take longer than 120 seconds to complete the injection, the
Injection
process.
Figure 110 shows the
Batch Injection Measurement
show the batch measurement procedure.
the analyzer:
message displays and instructs you to restart the injection
Figure 110: Batch Injection Flow
screen displayed as a flow diagram to
Failed