PART 5
MAINTENANCE
Annual routine maintenance
To secure section quality and functioning of the
STS, it is recommended that a routine
maintenance should be performed by a trained
service technician once a year.
Service Contract
Thermo Fisher Scientific Microm International
GmbH offers a service contract which
guarantees that your STS is always in perfect
condition. For more information, please contact
the nearest sales office.
Exchange of fuses:
The two mains fuses are installed on the rear
of the control unit beside the power switch.
To exchange the two fuses, unplug the unit.
Open the drawer (fig. 10.3) from the left side
by means of a flat screwdriver and pull it out.
Pull out the fuse holders and replace them with
new ones.
While the drawer (fig. 10.3) is open, it is
possible to turn the inserted voltage selector
and thus to change the power requirements of
the instrument. In doing so, please note that
the set power requirements of the instrument
correspond with the power supply voltage
being used.
For a power supply voltage between 220 and
230 V, set the voltage selector to 230 V.
Rating of fuses:
For power requirements 100 – 240 V:
- 2 pieces, T2.5 AH, 5x20 mm
4
MICROM
GmbH
69190 Walldorf, Germany
Type STS-II
Prototyp 5
__A /___V +10% __Hz
Fuse 2 x __AT
Made in Germany
1
Thermo Scientific Section Transfer System Microm STS
2
3
Fig. 10
Microm International GmbH
Part of Thermo Fisher Scientific
Im Steingrund 4 - 6
63303 Dreieich / Germany
387 712 - English
Caution:
Any shipping of the instrument requires original
packaging materials! Damages caused by
shipping with non-conform packaging are not
covered by the manufacturer warranty! Any
damage repairs resulting in non-conforming
package are fully charged to the sending party.
We
reserve
the
seriousness of damage NOT to repair.
To order original packaging materials, please
contact Thermo Fisher Scientific or your local,
by Thermo Scientific authorized, dealer.
Caution:
The user must care for a clean and safe
condition of the instrument when returning
it to an appropriate service provider.
Note:
If the original packaging is no longer available,
please contact your local Thermo Scientific
representation.