Thermo Scientific ERGOSTAR HM 200 Instruction Manual page 11

Section transfer system microm sts for rotary microtomes

Advertisement

2
OPERATING INSTRUCTIONS
2-1
Initial turn-on
Note:
The work sequence which is described
in this part must strictly be observed.
To use a rotary microtome together with a
Section
Transfer
following preparations are necessary:
 Pull off the integrated section waste tray of
the rotary microtome.
 Unplug the possibly connected foot pedal
and the power cord.
 Then tilt the rotary microtome backwards and
replace the shorter rubber feet with the
higher feet for correct height adjustment. (If
necessary, use appropriate pliers).
 Again bring the rotary microtome into working
position.
 Insert the disposable blade carrier TE (fig.
2.3) onto the dovetail guide of the rotary
microtome.
 Then evenly slide the STS housing (fig. 2.2)
from the front onto the rotary microtome.
Place the control cable, which is mounted on
the left front side of the STS housing, under
the microtome and lead it to the rear side.
Note:
The control cable ends in a multi-plug.
Insert this plug into the rear wall (fig.
3.1) of the control unit (fig. 4) and
secure it via the lateral screws.
Thermo Scientific Section Transfer System Microm STS
System
(STS),
the
Microm International GmbH
Part of Thermo Fisher Scientific
Im Steingrund 4 - 6
63303 Dreieich / Germany
387 712 - English
3
10
0
1
4
MICROM
GmbH
69190 Walldorf, Germany
Type STS-II
Prototyp 5
__A /___V +10% __Hz
Fuse 2 x __AT
Made in Germany
2
1
2
Fig. 2
3
Fig. 3
Fig. 4
11