Thermo Scientific ERGOSTAR HM 200 Instruction Manual page 26

Section transfer system microm sts for rotary microtomes

Advertisement

4-5
How to avoid malfunctions
Caution:
Turn on the STS only after having filled
the
system
demineralized water.
 To fill the system, only use distilled or
demineralized water to keep furring on the
transfer track at a minimum.
 Avoid pollution of the pump reservoir water,
to protect the pump from malfunctions.
 Select a suitable clearance angle. Set a
clearance angle adjustment of 5 – 15°
according to the facet angle.
 The blade should be free of paraffin.
 Make sure that no debris is between
detaching tongue and blade.
 Carefully wet the transfer track by using a
brush. If necessary, repeat process after
longer breaks.
Further information to avoid malfunctions can
be found in part 4-4 of the instruction manuals
of the respective rotary microtome.
26
Thermo Scientific Section Transfer System Microm STS
with
distilled
or
Microm International GmbH
Part of Thermo Fisher Scientific
Im Steingrund 4 - 6
63303 Dreieich / Germany
387 712 - English