 Via the lateral adjusting wheel increase the
pump capacity that there is an even water
flow on the transfer surface (fig. 8.4).
 The pump capacity is too high, in case water
flows down behind the blade on the
disposable blade carrier TE.
Note:
First,
thoroughly
transfer surface using a brush and
having selected a low pump capacity.
Only an even water flow on the transfer
surface leads to optimal operation processes
with the STS.
Note:
The transfer surface is moistened
enough when, after having activated
the detaching tongue (fig. 8.5) via the
detaching lever (fig. 8.6), the sections glide on
the water downwards on the transfer surface
with an even speed.
5
6
14