4.
Log in to the unit.
Tip
For user name and password, contact your local Sony Sales Office/Service Center.
(1) Enter user name and press the Enter key.
[pass:] appears.
(2) Enter password and press the Enter key.
[>] appears to allow you to enter a command.
Tip
Example of notation for commands input:
␣ : Space for one character
•
• ⏎ : Enter key
After each command has been entered, check the return value and then enter the next command or perform the
following procedure.
• Several seconds are required until "ok" appears in some cases.
• If "ng" appears, re-enter the command.
Note
At the end of the work, be sure to disconnect the unit from the PC by using the TeraTerm.
The TeraTerm can no longer be used.
1-4-7.
How to Check the Versions
Check the program, kernel, FPGA and CPLD versions from the PC.
Equipment required
• Personal computer (PC)
OS: Windows 7 or Windows8, Windows8.1
• LAN cable (cross cable)
Preparation
1.
Connect the unit to the PC with LAN cable.
2.
Setup the PC. (Refer to
Procedure
1.
Turn on the unit.
2.
Run the TeraTerm and perform the DgTerm command connection from the PC. (Refer to
Commands from the
PC".)
3.
Input the "ver " after "> ", and press the Enter key.
Version number is displayed.
Program: version x.xx.xx.x
Kernel: version x.xx
ATLANTIC3: version xx
CPLD_CPC: version xxx
CPLD_CPF: version x
FPGA_CORE: version xxx
CPLD_CPV: version x
FPGA_TC: version xxx
BVM-E171
"1-4-4. PC
Setting".)
"1-4-6. Connecting
1-12