Battery Management - Scaleo medical Shado 160 User Manual

Shower trolley
Table of Contents

Advertisement

Battery management

When the battery discharge threshold is reached, the user is alerted by a beep. If the battery does
not recharge quickly, the unit turns off and the raising/lowering functions are not active anymore.
A power reserve allows the Shado
showering of a patient enabling it to finish the transfer.
The AC plug is protected by a plastic cap. The cap must be removed for the charge, and put
back in place once the battery is fully charged.
Charging the battery
Use only the supplied cable to charge the shower trolley: charging a
battery with an unsuitable cable can cause the cable and/or battery
to overheat and damage all components, or cause a fire.
If the emergency stop button is pressed, the battery will not charge.
To charge the battery:
1. Connect the AC cable to the AC outlet,
2. The green «ON» light turns on,
3. The orange «CHARGE» light turns on, only if the battery is connected correctly.
It takes 8 to 10 hours to recharge 100% empty batteries.
After a charge cycle, do not forget to cut out the main/24VDC transformer from the main
circuit then from the Shado
guaranteed if the feeding cord is cut (even partially).
The batteries cannot be overcharged. However, to ensure better performance and longevity,
it is recommended to let them rest for a few minutes after charging and before use. It is also
important to fully charge the battery.
Always unplug the power cord after each charge, without pulling it to avoid damaging the
cord.
DOMA019_EN_v16_20221206_Shado® shower trolleys user manual
®
to be operated if the sound signal rings during the
®
connection current collector. The transformer is not
15

Advertisement

Table of Contents
loading

Table of Contents