Cleaning and disinfecting the circuit

"Cleaning" mode
It is recommended to proceed daily to the washing and the disin-
fecting of the liquid circuit of the unit.
The "Cleaning" mode allows for the washing and the disinfecting of
the liquid circuit in order to reduce the crystalline deposits and the number
of bacteria in the liquid circuit. A rinsing of the circuit is then required to
avoid a possible reaction caused by the ageing of the solution or by an
unwanted mixture of liquids.
All the bottles and handpiece cords used during treatment must
be present during the washing and the disinfecting of the liquid circuit
because the process is simultaneous for all these elements. Remove the
handpieces.
Both connectors must be cleaned.
Circuit washing cycle:
Prepare the bottle(s) with distilled or demineralized water,
Press the "Cleaning" mode selector,
Select the bottle(s) to use,
Lift the connector(s) and put it (them) above a sink,
Press the foot pedal to start the program. Three "beeps" and the
blinking of the control lamp of the "Cleaning" mode confirms the end
of the cycle.
Disinfecting cycle:
Prepare the bottle(s) with a compatible disinfecting solution (refer to
the list of compatible disinfectants at the beginning of this manual),
Select the bottle(s) to use,
38
FB-430_CN_rev_C_PIEZON MASTER 700.indd 38-39
FB-430_CN_rev_C_PIEZON MASTER 700.indd 38-39
Lift the connector(s) and put it (them) above a sink,
Press the foot pedal to start the program. Three "beeps" signal and
the blinking of the control lamp of the "Cleaning" mode confirms the
end of the cycle,
Then allow the disinfecting solution to act for at least 5 minutes
without intervene with the unit.
Rinsing cycle:
Prepare the bottle(s) with the sterile distilled water,
Select the bottle(s) to use,
Lift the connector(s) and put it (them) above a sink,
Press the foot pedal to start the program. Three "beeps" signal and
the blinking of the control lamp of the "Cleaning" mode confirms the
end of the cycle,
Press the "Cleaning" mode selector to exit this mode (this cycle must be
then restarted).
(Note: it is possible at any time to press the foot pedal to stop the program
in progress.)
The informative durations of the "Cleaning" mode cycles are defined as
follows (Note: one led corresponds to 10 seconds):
1 bottle
2 bottles
1 cord
30 seconds
1 minute
50 ml
2 x 50 ml
2 cords
1 minute
2 minutes
100 ml
2 x 100 ml
清洁和消毒管路
"清洁"模式
建议每天对装置的液体管路进行冲洗和消毒。
"清洁"模式允许对液体管路进行清洗和消毒,以减少液体管路中
的结晶物积聚和细菌数量。 之后,还需要冲洗管路,以避免溶液陈化或
液体发生不必要的混合引起反应。
清洗和消毒液体管路时必须对治疗过程中使用的所有专用水瓶和洁
牙手柄线进行处理,因为这一过程对于所有此类元件是同步的。 拆下洁
牙手柄。
两个连接器必须清洗。
管路清洗循环:
准备好装有蒸馏水或脱矿物质水的专用水瓶,
按"清洁"模式选择器,
选择要使用的专用水瓶,
抬起连接器并将其放在水槽上,
按下脚踏开关并启动程序。 会发出三声"峰鸣声","清洁"模式
控制灯会闪烁,确认清洁循环结束。
消毒循环:
准备好装有相容消毒液的专用水瓶(请参见本手册开头的相容消毒
剂清单),
选择要使用的专用水瓶,
抬起连接器并将其放在水槽上,
按下脚踏开关并启动程序。 会发出三声"蜂鸣声","清洁"模式
控制灯会闪烁,确认清洁循环结束,
然后允许消毒液作用 5 分钟,在此期间不要干扰装置。
冲洗循环:
准备好装有除菌蒸馏水的专用水瓶,
选择要使用的专用水瓶,
抬起连接器并将其放在水槽上,
按下脚踏开关并启动程序。 会发出三声"蜂鸣声","清洁"模式
指示灯会闪烁,确认清洁循环结束,
按"清洁"模式选择器退出此模式(之后必须重新启动该循环)。
(注:随时可以按脚踏开关停止正在执行的程序。)
"清洁"模式周期的信息持续时间定义如下(注意:一个 LED 对应 10
秒钟):
1 瓶
2 瓶
1 条线
30 秒
1 分钟
50 ml
2 x 50 ml
2 条线
1 分钟
2 分钟
100 ml
2 x 100 ml
4/7/2020 9:55:39 AM
4/7/2020 9:55:39 AM
39

Advertisement

Table of Contents
loading

Table of Contents