System Parameter Mode - Yamato PPC-200W Technical Manual

Hide thumbs Also See for PPC-200W:
Table of Contents

Advertisement

PPC-200W
II Set-up Procedures
II.C

System Parameter Mode

1. Enter Test Mode (see section II.A for details.)
2. Press
and
scale will indicate a keyword fol lowed by it's value.
3. To select the desired keyword, use the keys as follows:
- increase keyword by one
+
Note: Use the System Parameter List in Section IV.B to locate
the function to be changed and it's standard value.
Example: To select zero tracking time press
times.
4. User parameters are also set in the System Parameter Mode.
Press
to cycle through system keywords 30 through 40.
After system keyword 40, user keyword 01 is displayed. Press ing
will continue to cycle through user keywords 01 through 09
and then return to system keyword 30.
Note: Use the User Parameter List in Section IV.A to locate the
function to be changed and it's standard value.
5. To change a parameter value, use the keys as follows:
- increase value by one
+
- decrease value by one
+
to enter System Parameter Mode. The
- decrease keyword by one
- increase value by ten
- decrease value by ten
Technical Manual
Key word
eight
3
Parameter Value
YCO-1049-02-03