Calibration of current input (0 to 20mA)
In a 0 to 20mA current input calibration connect a current source to the
channel to be calibrated (see the terminals corresponding to each channel in table 4
on section 2.3.3 - Milliampere Input or tables 6 and 7 on section 2.3.5 - Connection
for Indicators with different input types). It is required 6 current references listed in
table 12.
Table 12 - References for current input calibration
Calibration of voltage input (0 to 5V)
In a 0 to 5V voltage input calibration connect a voltage source to the channel
to be calibrated (see the terminals corresponding to each channel in table 5 on
section 2.3.4 - Voltage Input or tables 6 and 7 on section 2.3.5 - Connection for
Indicators with different input types). It is required 6 voltage references listed in table
13.
Table 13 - References for voltage input calibration
Reference
0.000 mA
4.000 mA
8.000 mA
12.000 mA
16.000 mA
20.000 mA
Reference
0.0000V
1.0000V
2.0000V
3.0000V
4.0000V
5.0000V
DMY - 2015
Mnemonic
C. 0nA
C. 4nA
C. 8nA
C.12nA
C.16nA
C.20nA
Mnemonic
C. 0V
C. 1V
C. 2V
C. 3V
C. 4V
C. 5V