Presys DMY-2015 Technical Manual page 28

Digital multi-point indicator
Table of Contents

Advertisement

Presys  Instruments
When selecting a linear sensor one must configure its scale (SCALE option).
Define two points P1(Lim Low, Eng Low) and P2(Lim High, Eng High), as illustrated
in figure 14. Lim Low represents the value of the electrical signal given in % of full
scale associated to the Eng Low indication on the display, and Lim High corresponds
to the value of the electrical signal given in % of full scale associated to the Eng High
indication on the display.
SQRT - allows presenting on the display the squared root of the linear input
signal. The Cut-Off parameter given in % of the input signal makes input values
below (Lim Low + Cut Off) be shown as Lim Low. See figure 15.
(*) % OF FULL SCALE OF INPUT SIGNAL
Fig. 14 - Linear input configuration
Fig. 15 - Input signal square root
DMY - 2015