System Indicators; Advantage Controller Front Panel (Top Portion) - ConMed Advantage Instruction Manual

Table of Contents

Advertisement

PORT 1 PORT 2
11
12
1.6

System Indicators

1.6.1
Advantage
Controller Front
Panel (Top Portion)
1.
Footswitch Receptacle — Accepts
the footswitch cord connector.
2.
APEX
Handpiece Receptacle
(Port 1) — Directly accepts the
handpiece cord connector of any
handpiece (C9820, C9824, C9828, C9840).
3.
Handpiece Receptacle (Port 2)
Directly accepts the handpiece
cord connector of the following
handpieces; all Group 1
Handpieces (E9005, E9015, and CoolFlex);
all Group 2 Handpieces (5020-021, -022,
-023, -024, -025, -026, -027); most Group 3
Handpieces (D9820, D9920, MC9820,
D9824, D9924, MC9828, and MC9840);
and the
PRO2300E and
PowerPro
PRO6125 Oscillator Saws, PRO6100
PORT 3
APEX
13
Electric II Modular Handpiece, and
PRO6135 Reciprocator Saw. However,
with the addition of a handpiece
adapter - the
APEX
Mini-Driver
Adapter (K501) - this
receptacle will also accept and operate the
following shaver handpieces (C9820,
C9824, C9828, C9840) or the
Handpiece (K500).
NOTES:
All cord receptacles and connectors are
keyed to prevent incorrect insertion.
Likewise, receptacles and connectors are
color-coded to easily recognize where to
connect each cord (footswitch receptacle
and connector - yellow, handpiece
receptacles and connectors - red or grey).
Only one handpiece can be connected at
a time in either port of the top section. If
a handpiece is connected while another
handpiece is already connected,
"CHECK HANDPIECE" will display.
BOTH handpieces must be removed to
clear the fault condition.
9
Adapter (E9320) or the
Mini-Driver