3.4.4
Low Speed Drill (5020-026)
Operation
PRECAUTION: Make sure the head
attachment matches the function
selected. Improper head attachment
selection may result in erroneous speed or
torque.
NOTES:
1.
The Low Speed Drill can only be
operated by the footswitch. Therefore, it
can only be operated on the
controller or Port 2 of the
controller (Top Portion of controller).
2.
Always operate the Low Speed Drill
with the appropriate attachment,
either the Dental Implant Drill Head
(5020-034) or the Implant Screw/Tap
Head (5020-035).
3.
The Low Speed Drill accepts standard
dental latch burs or
C
the 5088 series.
4.
When the Low Speed Drill is set to
"SCREW" or "TAP" mode, the speed is
not variable in the reverse direction.
5.
If 25 in. oz. is exceeded while using this
handpiece, the handpiece will stop and
"MAXIMUM
The handpiece can only be reset by
secondary activation of the footswitch.
E9000
Advantage
Hall Surgical
burs of
TORQUE" will display.
1.
To connect an attachment:
(a) Place the "SAFE/RUN" Slide in the
"SAFE" position. The controller will
display the current operating mode
("SCREW" "TAP"
ON", and the set flow rate
"SAFE
("OFF" will display if irrigation is not
being used).
(b) Slide the appropriate attachment over
the front end of the handpiece and seat
completely.
80
"DRILL"),