Output Power - ConMed Sabre Genesis Service Manual

Electrosurgical generator
Table of Contents

Advertisement

4.5.3

Output Power

1) Equipment Requirements:
a) Monopolar Footswitch
b) Bipolar Footswitch
c) Commercial ESU Tester (e.g. Fluke 454A
or equivalent) with non-inductive 50 load
for bipolar modes and a non-inductive
500 ohm load for monopolar modes.
Note: Micro Bipolar is particularly sensitive to
the load resistance. A 50 ohm load should be
used for checking power to obtain the best
Mode
Load (ohms) Power Setting Watts (min) Watts (max)
Pure
500
500
500
Blend
500
500
500
Table 4.2 Monopolar Coag Mode RF Output Power Accuracy
Mode
Load (ohms) Power Setting Watts (min) Watts (max)
Standard
500
500
500
4) Disconnect the ESU tester from the unit.
5) Use test leads to connect the ESU tester to
the Bipolar Accessory outputs.
Mode
Load (ohms) Power Setting Watts (min) Watts (max)
Macro
50
50
4-2
Power Accuracy
20
17
100
90
200
180
20
17
100
90
200
180
20
17
50
45
80
72
Table 4.3 Bipolar Mode RF Output Power Accuracy
20
17
50
45
results.
2) Use test leads to connect the ESU tester to
the unit's return electrode output and the
footswitch controlled active output. Set the
Load resistance per mode as indicated in
Tables 4.1 and 4.2.
3) Perform the monopolar power tests indicated
in Tables 4.1 and 4.2. The acceptance range is
given in both Watts and Amps to accommo-
date available test equipment. It is not neces-
sary to test for both power and current.
Table 4.1 Monopolar Cut Mode RF Output
Amps (min)
23
0.184
110
0.424
220
0.600
23
0.184
110
0.424
220
0.600
Amps (min)
23
0.184
55
0.300
88
0.379
6) Perform the bipolar power tests indicated in
Table 4.3. This table only provides the mini-
mum number of points to be tested.
Amps (min)
23
0.238
55
0.387
Amps (max)
0.214
0.469
0.663
0.214
0.469
0.663
Amps (max)
0.214
0.332
0.420
Amps (max)
0.277
0.428