2
Safety
First aid in case of shortness of
First aid in case of frostbite
46

Medical emergency

2.4.2
 WARNING
Medical emergency during MR measurements!
Risk of death to patients
Terminate the measurement immediately.
Remove patients from the examination room for treatment
unless it is certain that the medical equipment required is
appropriate for use inside an MR room.
Do not store or operate oxygen bottles, defibrillators, or other
auxiliary tools for resuscitation in the examination room.
 CAUTION
Squeeze bulb is defective!
Risk of injury to patient because emergencies cannot be
communicated
Check the functionality of the squeeze bulb daily.
2.4.3

Coolant accidents

During a quench, a person might become unconscious due to severe
breath
shortness of breath:
1
Remove unconscious persons immediately from the examination
room.
2
Start adequate first aid measures and contact a physician
immediately.
Direct contact with subzero liquids, gases, and surfaces (e.g. pipes)
may lead to frostbite. The eyes and mucous membranes are
especially vulnerable.
MR System | Operator Manual
Print No. MR-01501G.621.02.02.02