Siemens MAGNETOM Avanto Operator's Manual page 75

Hide thumbs Also See for MAGNETOM Avanto:
Table of Contents

Advertisement

Stopping the patient table
movement
Terminating the measurement
sequence
Acknowledging the patient
Listening to physiological
monitoring signals
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
Print No. MR-01501G.621.02.02.02
– or –
Go to the control unit and turn the volume control fully to the left
(counter-clockwise).
Press the Table Stop button.
The table movement is stopped immediately.
For interventions outside the magnet, movements of the patient
table can be stopped immediately by pressing the Table Stop
button. The measurement sequence is terminated. After
completion of the intervention, the control unit at the magnet is
enabled again.
Press the Table Stop button very shortly one after the other.
The measurement is terminated.
The patient alert is activated. An acoustic signal sounds at the
alert
intercom. The Squeeze bulb/announcement active LED flashes.
Press the Speak button.
The signal stops sounding. The patient alert is switched off. The
transmission from the examination room is activated. It is now
possible to speak with the patient.
The operating personnel may use the intercom to listen to the trigger
pulses of the physiological monitoring signals.
The volume of the trigger pulses can be set to three different levels:
Low
Average
High
Use the Trigger signal volume control to set the trigger pulse
volume.
The trigger pulses are transmitted at the desired volume.
MR system components
3
75