(1) Magnet Stop, example individual switch: "Press to remove field.
Emergency use only."
(2) Magnet Stop on the different versions of the alarm box: "Press
to remove field. Emergency use only."
After the Magnet Stop switch has been pressed, an alarm is triggered
at the alarm box. The MAG STOP or WARNING LED will light up and
an alarm signal will sound.
As a rule, Siemens Service must be called following a quench.
The magnet must only be put back into operation by Siemens
Service personnel.
 WARNING
Formation of droplets due to condensation during quenching!
Personal injury (for example, frostbite)
Risk of fire
Do not touch the exhaust line / quench pipe.
Do not stand under the exhaust line.
Avoid open flames and do not smoke.
Safety