Siemens MAGNETOM Avanto Operator's Manual page 23

Hide thumbs Also See for MAGNETOM Avanto:
Table of Contents

Advertisement

MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
Print No. MR-01501G.621.02.02.02
 CAUTION
The patient is wearing electrically conductive material! Incorrect
patient positioning!
Formation of electric current loops
Peripheral nerve stimulation of the patient
Ensure that the patient does not wear clothing that is wet or
dampened by perspiration.
Ensure that the patient is free of metallic rings, chains, or
electrically conductive materials worked into items of clothing
(for example, brassiere support wires, metallic appliqués or
woven metallic yarns).
Always position the patient so that the patient's arms are
aligned with the torso and ensure that hands, arms, and legs
do not touch (minimum distance: 5 mm).
Ensure that the minimum distance of 5 mm is maintained
between patient and tunnel covering.
To ensure this distance, use positioning aids, e.g. blankets
made of linen, cotton, or paper, or dry material that is
permeable to air.
Ensure sufficient ventilation.
 CAUTION
Heating up/ignition of synthetic blankets via the RF field during
the measurement!
Patient burns
Use only covers made of paper, cotton or linen.
Safety
2
23