ECG and respiratory sensor
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
Print No. MR-01501G.621.02.02.02
The physiological signals are acquired with receptors - ECG
electrodes, respiratory cushion and pulse sensor - directly at the
patient via the PERU (ECG, respiration) and PPU (pulse). The
measured data can be viewed at the PMU display in the examination
room and in the Physiological Display dialog box at the syngo
Acquisition Workplace.
Only one PERU or PPU can be located in the examination room!
Two ECG and respiratory sensors in the examination room
interfere with one another in signal transmission. It is not
possible to determine the results.
All components of the MR system mentioned here must only be
used to control MR measurement sequences. They are not
approved as a patient monitoring system!
The wireless PERU simultaneously acquires two ECG channels as well
(PERU)
as the respiratory channel of the patient.
PERU version 1 (A) and PERU version 2 (B)
(1) ECG leads with clips
(2) Transmitter unit
(3) Control LEDs
(4) Plug for respiration cushion
(5) Connector, charger
The ECG electrodes and respiratory cushion are connected to the
PERU.
Physiological imaging
4
87