If the audible/visual preventive maintenance alert sounds, complete the
Preventive Maintenance Form before resetting the lift counter. Instructions to
reset the lift counter are detailed in the AP-450 Service/Spare Parts Manual.
Annually
(To be completed by a lift technician)
This section to be only completed by a qualified service technician as
authorized by Handicare.
Annually, the AP-450 system must be inspected. Be sure to documents this service history.
Complete the inspection as noted in the "Monthly" section above.
Complete the preventive maintenance procedure as outlined in the AP-450 Technical
Manual.

Recurrent Performance and Safety Testing

The following table highlights AP-450's Essential Performance Parameters, with their subsequent testing
method and frequency. If your AP-450 does not meet the following criteria, contact your local Handicare
authorized dealer for service.
Performance Criteria
The AP-450 should be
capable of lifting 450 lb
(204 kg).
The AP450 will operate as
intended throughout its
service life, provided all
recommended
maintenance and usage
recommendations are
followed.
The AP450 should be
capable of working in
conjunction and
compatibility with
Prism/Handicare Track,
trollies and accessories
(Carry bars and Slings).
The AP450 should have
"Free Fall" protection
system.
The AP-450 should have
upper and lower limits.
Recommended Check
Load test at full load
(450lb/204kg)
1. Load test at full load
(450lb/204kg)
2. Pre-lifting inspection by
user care giver.
3. Preventive maintenance
by trained tech.
1. Load test at full load (450
lb)
2. Pre-lifting inspection by
user care giver.
3. Preventive maintenance
by trained tech.
Preventive maintenance by
trained tech.
Preventive maintenance by
trained tech.
Annual
1. Annual.
2. Every use
3. Semi-annually or annually, per usage
frequency.
See "General Inspection and Preventive
Maintenance" on page 17.
1. Annual.
2. Every use
3. Semi-annually or annually, per usage
frequency.
See "General Inspection and Preventive
Maintenance" on page 17.
Semi-annually or annually, per usage
frequency. See "General Inspection and
Preventive Maintenance" on page 17.
Semi-annually or annually, per usage
frequency. See "General Inspection and
Preventive Maintenance" on page 17.