Handicare AP-300 Owner's Manual page 20

Table of Contents

Advertisement

Performance Criteria
The AP-450 strap should
have a minimum safety
factor of 6 in terms of
break load.
The AP-450 should have
visual "Error" status
display.
The AP-450 hand control
locking mechanism used
as connections should be
"Self-Locking".
Recommended Check
1. Load test at full load (450
lb)
2. Pre-lifting inspection by
user care giver.
3. Preventive maintenance
by trained tech.
Preventive maintenance by
trained tech.
Preventive maintenance by
trained tech.
Annual
Every use
Semi-annually or annually, per usage
frequency
See "General Inspection and Preventive
Maintenance" on page 17.
Semi-annually or annually, per usage
frequency.
See "General Inspection and Preventive
Maintenance" on page 17.
Semi-annually or annually, per usage
frequency.
See "General Inspection and Preventive
Maintenance" on page 17.
20
Frequency

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ap-450Ap seriesAp300

Table of Contents