CardioQuip MCH-1000 Operators & Service Manual page 40

Cardiopulmonary bypass temperature controller
Hide thumbs Also See for MCH-1000:
Table of Contents

Advertisement

CAUTION: Do not plug the Refrigeration Module into the same electrical circuit
as the MCH-1000(i). The Refrigeration Module draws approximately 6 Amps of
power, so it may be able to share a circuit with other equipment. The MCH-1000(i)
requires a separate dedicated 20 Amp circuit (see Initial Operation Check).
CAUTION: Do not damage the chiller plates or temperature probe. Bending or
twisting the plates or probe can cause premature failure of the Refrigeration Module.
CAUTION: Always keep the Refrigeration Module upright. Setting the module on
its top or sides can damage the compressor.
CAUTION: The water tank should be at least 1/2 full of sterile or 0.22-micron
filtered water when using the Refrigeration Module. If the water depth in the tank
is too low, the Refrigeration Module will be unable to produce enough ice for normal
device operation, and compressor damage could result. Fill the tank to the top edge
of the chiller plates with sterile or 0.22-micron filtered water.
CAUTION: Do not block the air vents around the sides of the Refrigeration
Module. Inadequate airflow will cause overheating and premature failure of the
Refrigeration Module.
CAUTION: Do not operate the Refrigeration Module with the hinged cover open.
An open cover inhibits cooling airflow into the unit, and may allow aerosolization of
water-borne pathogens that could be present in the water tank.
38
OPTIONAL COOLING MODULES | Refrigeration Module