SETTING THE TEMPERATURE: 3 WAYS

The set temperature can be adjusted from the STANDBY, RUN, or
PRIMING screens. The MCH offers three different methods of adjusting
the set temperature:

Tenth-Degree Adjustment

To set the desired temperature, from 0˚C to 42˚C in tenth-degree
increments, touch the
these buttons for more than one second will cause the temperature
setting to change more rapidly.

Manual Input

Touching the dark blue SET POINT temperature field will
cause a numeric keypad to appear onscreen, allowing direct
input of the desired temperature. After entering the desired
temperature, touch the
to cancel. The MCH will not allow the set temperature
to be higher than 42˚C or lower than 0˚C.

Presets

The set temperature may also be changed using the
button. This brings up a set of buttons with preset
temperatures. Touching one of these buttons changes the
temperature Set Point to the preset temperature on the
button.
To edit the preset temperatures, touch the
The temperature buttons will turn yellow. Touch a button to
edit the temperature on that button. A numeric keypad will
appear onscreen, allowing direct input of the desired
temperature preset. After entering the desired temperature
preset, touch the
(Escape) to cancel.
When you are finished editing the presets, touch the
button to return to the preset selection screen. The
temperature setting and temperature presets are stored in
non-volatile memory, and will be retained until changed.
or
temperature control button. Holding
button to confirm, or touch
(Enter) button to confirm, or touch
Presets
button.