System Drain; Tank Drain - CardioQuip MCH-1000 Operators & Service Manual

Cardiopulmonary bypass temperature controller
Hide thumbs Also See for MCH-1000:
Table of Contents

Advertisement

System Drain

The MCH-1000(i) system drain valve is located on the underside of the unit, and allows the
operator to drain all or part of the water from the system. To open the valve, move the valve
lever down to the vertical position. To close the valve, move the lever to the horizontal
position.
The MCH-1000(m) system drain connector is located
on the rear of the unit, below the main hose connectors,
and operates in the same manner as the blanket hose
connectors on the MCH-1000(i).

Tank Drain

The tank drain valve is located on the front panel of the MCH-1000(i), and allows the operator
to drain water from the water tank without draining the entire system. This may be useful e.g.
to make room in the tank for more ice made with sterile or 0.22-micron filtered water during a
procedure. To open the valve, turn the valve lever so that it is parallel with the valve body. To
close the valve, turn the lever to that it is perpendicular to the valve body (see photos below).
A standard garden hose may be attached to transfer the water to a remote drain.
System Drain Open
Tank Drain Open
System Drain Closed
Tank Drain Closed
CONNECTORS, VALVES, & HOSES | System Drain
MCH-1000(m)
System Drain
29