Foot Board Maintenance Procedures; Foot End Control Membrane Replacement - Stryker FL26E Maintenance Manual

Electric med/surg bed
Table of Contents

Advertisement

4.2 FOOT BOARD MAINTENANCE PROCEDURES

FOOT END CONTROL MEMBRANE REPLACEMENT

Required Tool:
Small Slotted Head Screwdriver
Procedure:
1. Fully raise the bed and apply the brakes.
2. Unplug the bed power cord from the wall outlet.
3. Using a small slotted head screwdriver, lift and remove the self-sticking membrane. Proceed
gently when inserting the screwdriver under the membrane to avoid scratching the molded
board.
4. Pull gently on the membrane cable to free the membrane connector from the foot board
opening.
5. Disconnect the membrane and connect the new membrane. Slide the membrane tail back in
the foot board opening and stick the membrane on the foot board.
N
O
T
E
N
O
T
E
Be careful not to stick the cable on the back of the membrane.
6. Test all foot board controls for proper operation before returning the bed to service.