2.7 FOOT BOARD CONTROL PANEL GUIDE
SIDERAIL CONTROL
LOCKOUT
SWITCHES
A1
A2
CONTOUR
F
WHEN ON
•
Lockout Switches (A1 to A3)
These three switches enable the selective lock out of the bed functions available to patient
and nursing staff through the siderail control panels (inner and outer).
A1
: Push to inhibit access to Fowler adjustment from both inner and outer siderail control
:
panels. The associated LED will go off.
A2: Push to inhibit access to Knee Gatch adjustment from both inner and outer siderail
control panels. The associated LED will go off.
A3: Push to inhibit access to bed Hi-Lo control from the siderail outer control panels. The
associated LED will go off.
N
O
T
E
N
O
T
E
When a lockout switch is used to inhibit a specific function of the siderail control panels, the
function is inhibited on both sides of the siderail control panel. The foot board controls are not
affected by the lockout switches.
Lockout settings are automatically saved in the event of a power failure and restored following
resumption of power.
•
Fowler Control Switches (B1 and B2)
These two switches enable the adjustment of the Fowler angle.
B1: Push to raise Fowler.
B2: Push to lower Fowler.
•
Knee Gatch Control Switches (C1 and C2)
These two switches enable the adjustment of the Knee Gatch angle.
C1:Push to raise Knee Gatch.
C2: Push to lower Knee Gatch.
SLEEP SURFACE POSITION SWITCHES
HEAD
A3
B1
B2
AUTO CONTOUR
PICTOGRAM (OPTIONAL)
Operation Guide Chapter 2
THIGH
HI-LO
C1
C2
D1
D2
TRENDELENBURG
E
SWITCH
15