Siderail Control Membrane Replacement; Siderail Outer Panel Control Board Replacement - Stryker FL25E Maintenance Manual

Electric med/surg bed
Table of Contents

Advertisement

Maintenance Manual

SIDERAIL CONTROL MEMBRANE REPLACEMENT

Required Tools:
Small Slotted Head Screwdriver
Procedure:
D
A
C
B
1. Run the bed fully up and apply the brakes. Raise the siderail needing repair.
2. Unplug the bed power cord from the wall outlet.
3. Using a small slotted head screwdriver, lift and remove the self-sticking membrane (A) of the
inner or outer control panel needing to be replaced. Proceed gently when inserting the
screwdriver under the membrane to avoid scratching the rail.
4. Install the new membrane. Be sure to position the membrane properly over the board
switches.
5. Test all controls of the control panel for proper operation before returning the bed to service.

SIDERAIL OUTER PANEL CONTROL BOARD REPLACEMENT

Required Tools:
Small Slotted Head Screwdriver
Procedure:
N
O
T
E
N
O
T
E
Unless otherwise indicated, refer to figure 4.1C above for the illustration of this procedure
reference points.
Run the bed fully up and apply the brakes. Raise the siderail needing repair.
1.
Unplug the bed power cord from the wall outlet.
2.
Using a small slotted head screwdriver, lift and remove the following self-sticking parts: the
3.
two screw-covers (B), the "Lift to release siderail" label (C) and the control panel membrane
(A) to expose the screws holding the siderail cover. Proceed gently when inserting the
screwdriver under the labels to avoid scratching the molded rail.
22
K
I
F
G
H
E
Figure 4.1C
Phillips Screwdriver
J

Advertisement

Table of Contents
loading