8
Dismantling / Mounting in Case of Service
Figure 19: Version with 5-pole plug cou-
pling
Figure 20: Version with 7/9-pole plug cou-
pling
2.
Position the pivot with 3-pole plug coupling 6, keeping it straightly aligned,
and establish the assembly connection 5 by gently pushing towards the
LCH spring arm 6.
Only for version with 5-pole plug coupling
(See "Figure 19")
1.
Grease the pivot groove and the pivot 6 using a suitable anti-friction bearing grease
(e.g. Optimol).
NOTE – Plug positioning through the grooves
The 5-pole plug coupling 6 has a narrow groove 7 and a wide groove 8.
To establish the plug connection 5, the grooves 7/8 must fit into the corresponding
grooves of the counter piece.
2.
Position the pivot with 5-pole plug coupling 6, keeping it straightly aligned,
and establish the assembly connection 5 by gently pushing towards the
LCH spring arm 6.
Only for version with 7/9-pole plug coupling
(See "Figure 20")
Connecting the plug
1.
Connect the 6-pole plug 7.
2.
Screw 3 plug adapters 8 into each other in accordance with the colour code.
3.
Gently push the 6-pole plug 7 and the plug adapters 8 into the LCH spring arm
6.
Mounting the end device
4.
Grease the pivot groove and the pivot 6 using a suitable anti-friction bearing grease
(e.g. Optimol).
5.
Insert the pivot 6 into the LCH spring arm 6, keeping it precisely axially aligned.
93