Instructions for use/Technical description
ELAN 4 electro drill GA844
Document No.: TA014436 - Version: 1.0 - Document ID: SOP-AIC-5002244 Date/Time Printed/Viewed: 2022-04-13 16:01 (CET)
®
Effective

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Aesculap ELAN 4 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Braun Aesculap ELAN 4

  • Page 1 Effective Aesculap ® Aesculap Power Systems Instructions for use/Technical description ELAN 4 electro drill GA844 Document No.: TA014436 - Version: 1.0 - Document ID: SOP-AIC-5002244 Date/Time Printed/Viewed: 2022-04-13 16:01 (CET)
  • Page 2 Effective Document No.: TA014436 - Version: 1.0 - Document ID: SOP-AIC-5002244 Date/Time Printed/Viewed: 2022-04-13 16:01 (CET)
  • Page 3 Effective Document No.: TA014436 - Version: 1.0 - Document ID: SOP-AIC-5002244 Date/Time Printed/Viewed: 2022-04-13 16:01 (CET)
  • Page 4 Effective Aesculap ® ELAN 4 electro drill GA844 Legend Reverse button clockwise/counterclockwise rotation + oscillation mode switch 1 Product (ELAN 4 electro drill) 2 Rotational speed control button 3 Safety catch 4 Screw handle Motor speed control 5 Cable 6 Connector for control unit 7 Knob for left-rotation 8 Coupling Manufacturer’s batch designation...
  • Page 5: Table Of Contents

    These instructions for use apply to the ELAN 4 electro drill tool GA844. ■ Max. temperature 48 °C ► For article specific instructions for use and material compatibility and Saw operation with GB892R: lifetime information, see B. Braun eIFU at eifu.bbraun.com ■ 30 second application, 60 second pause ■ Warnings 5 repetitions ■...
  • Page 6 Effective ► Electrical systems generally heat up during continual operation. It is Ensure that the product and its accessories are operated and used only advised to give the system a break after use to cool down, as listed in the by persons with the requisite training, knowledge, or experience.
  • Page 7: Connecting The Accessories

    ► Check that the tool is coupled properly. To do this, pull on the tool. ► Please address your B. Braun/Aesculap partner or Aesculap Technical ► For GB891R: check that the tool attachment has locked into place. To Service with any inquiries in this respect; for a contact address, see do this, turn the tool attachment.
  • Page 8: Safe Operation

    Effective ► Ensure that the tool blades are not mechanically damaged. 6.3.1 Normal Operation ► Product operating in right rotation mode: Release product for use (ON position). ► ► Activate the knob for speed control 2. Run the product shortly at maximum speed in right and left rotation. ►...
  • Page 9: General Information

    Effective Note Preparation before cleaning If there is no final sterilization, then a virucidal disinfectant must be used. ► Prior to first mechanical cleaning/disinfection process: Install the Aesculap ECCOS holder system in a suitable tray. Note ► Insert products in the correct position into the Aesculap ECCOS holder For up-to-date information about reprocessing and material compatibility, system, see Fig.
  • Page 10: Manual Cleaning With Wipe Disinfection

    Effective Manual cleaning with wipe disinfection Phase Step Conc. Water quality Chemicals [°C/°F] [min] RT (cold) ≥2 Pre-cleaning Until visually clean RT (cold) ≥2 Cleaning with enzyme D–W pH-neutral* solution ≥5 Intermediate rinse – Drying Wipe disinfection >1 Meliseptol HBV wipes 50 % Propan-1-ol Final rinse RT (cold) 0.5 FD–W...
  • Page 11: Mechanical Alkaline Cleaning And Thermal Disinfection

    Effective 7.8.2 Mechanical alkaline cleaning and thermal disinfection Machine type: single-chamber cleaning/disinfection device without ultrasound Phase Step Water Chemistry/Note [°C/°F] [min] quality Pre-rinse <25/77 D–W ■ Cleaning 55/131 FD–W Concentrate, alkaline: – pH ~ 13 – <5 % anionic surfactant ■...
  • Page 12: Storage

    Modifications carried out on medical technical equipment may result in GA844 ELAN 4 electro drill loss of guarantee/warranty rights and forfeiture of applicable licenses. ► For service and repairs, please contact your national B. Braun/Aesculap agency. 12.2 Performance data, information about standards Service addresses Max. power approx.
  • Page 13: Short-Time Interval Operation

    30 second application, 60 second pause ■ 4 repetitions ► Detailed information concerning the disposal of the product is avail- ■ able through your national B. Braun/Aesculap agency, see Technical 30 min cooling time ■ Service. Max. Temperature 48 °C Saw operation with GB892R: ■...
  • Page 14 Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com ® Aesculap – a B. Braun brand SOP-AIC-5002244 (TA014436) 2019-09 Rev. 1 Document No.: TA014436 - Version: 1.0 - Document ID: SOP-AIC-5002244 Date/Time Printed/Viewed: 2022-04-13 16:01 (CET)