Subscanning Unit Controller Assembly - Siemens FUJI-Reader CR-IR347 Service Manual

Hide thumbs Also See for FUJI-Reader CR-IR347:
Table of Contents

Advertisement

10.18 Subscanning Unit Controller Assembly

CAUTIONS
• When servicing a PC board, be sure to wear a wristband to ground your body. If you do
not ground your body, the buildup of static electricity on your body may damage
electronic components on the PC board.
• Before removing a cover, be sure to turn OFF the high-voltage switch. Removing a
cover with the machine turned ON causes photomultiplier damage.
I Removal procedure
(1) Remove the following covers:
• Lower front cover
• Lower right-hand side cover
• Right-hand inner cover
"3. Removing and Installing the Covers"
009-058-00