I Check/adjustment procedure
• Check that the SZ1 connector cable is properly routed.
N INSTRUCTION N
After the light-collecting mirror is replaced with a new one, make shading/sensitivity
corrections.
• Check for image problems.
"9.3 Checking for Image Problems"
009-058-00
10.20.2000 FM2732 (2)
"9.2 Shading and Sensitivity Corrections"
CR-IR347 Service Manual
MC - 325