3-3. Connection Cable Installation

• Connect the connection cable to the left or right of the sensor head, selecting left or right depending on
the installation location.
• A water-proof cap is fi tted to both ends of the sensor head. Remove the water-proof cap from the end
where the cable is to be connected and connect the cable. (Water-proof cap hex. socket-head opposite
side 5 mm)
• Use a tightening torque of 1.0 N·m for the connector that is used for the connection with the sensor head.
Notes
• Do not remove the water-proof cap from the end that is not connected. If the cap comes off , re-install it
securely. Using the scale unit without the cap may result in damage to the sensor head.
• If an additional tightness of cable connections is required, it is possible to use a thread locking adhesives
to the threaded area of the connector, i.e. Loctite 242 or its equivalent.
Tightening torque:
1.0 N·m
Note
When the connector has been connected with a torque of 1.0 N·m, check to see that the O-ring of the
connector is correctly placed inside the sensor head as indicated below.
SQ57-Z Series
Connector
14
(E)
2
3
1
Water-proof cap
Fig. 3-14
Connector
Sensor head
Fig. 3-15
Sensor head
Connector
Sensor head
O-ring
SQ57-Z Series, CH22 Series