3-3. Connection Cable Installation

• Connect the connection cable to the left or right of the sensor head, selecting left or right depending on
the installation location.
• A water-proof cap is fi tted to both ends of the sensor head. Remove the water-proof cap from the end
where the cable is to be connected and connect the cable. (Water-proof cap hex. socket-head opposite
side 5 mm)
• Use a tightening torque of 1.0 N·m for the connector that is used for the connection with the sensor head.
• Apply a thread-locking agent to the threaded area of the connector used for connection in order to
prevent looseness. (Loctite 242 or equivalent)
Notes
Do not remove the water-proof cap from the end that is not connected. If the cap comes off , re-install
it securely. Using the scale unit without the cap may result in damage to the sensor head.
Thread-locking agent
Tightening torque:
1.0 N·m
Water-proof cap
14
(E)
2
3
1
Fig. 3-14
SQ57-Z Series, CH22 Series