Magnescale SmartSCALE SQ57 Instruction Manual page 44

Table of Contents

Advertisement

Available languages

Available languages

Example 2: Installation when a stop surface of the scale and the sensor head is not made with
the bracket.
1. Adjust the parallelism of the scale bracket and sensor head bracket with respect to the machine guide,
and then secure in place.
Adjust the height and parallelism of the sensor head bracket with respect to the scale bracket, and then
secure in place.
Sensor head bracket
2. Check the label attached to the scale to make sure that the serial No. matches that of the sensor head.
(See Fig. 3-6.)
3. Remove the label. (See Fig. 3-7.)
4. Adjust the parallelism with respect to the machine guide while measuring surface A with a dial gauge,
and tighten the set screws.
<Measurement method>
Measure near the positions at both ends of the eff ective length on surface A.
12
(E)
0.05 MG
Fig. 3-11
0.2
MG
A side (installation guide)
Fig. 3-12
 
77
±0.5
Sensor head
bracket
Scale bracket
Unit: mm
Unit: mm
SQ57-Z Series, CH22 Series

Advertisement

Table of Contents
loading

Table of Contents