Step
Description
5.1 Unlock the columns by pushing the
5
latch button away from the columns.
5.2 While holding the latch button open,
slide column assembly out of the
device by lifting and pulling on the
metal pull handle.
6
6.1 Remove the columns completely from
the device by pulling outward on the
metal pull handle.
6.2 Both columns are removed as one
piece.
7.1 To install new columns, first remove
7
the four (4) dust caps from the new
columns.
7.2 Make sure there is no dust or debris
where the dust caps were located.
8.1 Insert the new columns into the device
8
immediately after removing the dust
caps.
8.2 Push the columns until the latch makes
an audible click and returns to the
closed position.
8.3 Push and fold metal pull handle flush to
bottom of columns.
DO NOT: leave the column ends exposed.
IMPORTANT: You need to notify the device that you have replaced the columns. This can be done through the
device itself or through the Inogen Connect App.
©2022 Inogen, Inc. All rights reserved.