8.5.3 LOW PRIORITY ALARM (LEVEL 2)
The following low priority alarms are accompanied by one beep and a solid yellow light.
Display Icon
8.5.4 LOW PRIORITY ALARM (LEVEL 3)
The following low priority alarms are accompanied by two beeps and a solid yellow light.
Display Icon
©2022 Inogen, Inc. All rights reserved.
Description
Replace columns
Column replacement is required
within 30 days.
Extended start up
Oxygen concentration is <87% two
minutes after the device's start up
sequence and at least 10 breaths
have been detected within the last
minute.
Description
Battery low, attach plug
Battery power is low with less than
10 minutes remaining.
Oxygen low
The concentrator has been produc-
ing oxygen at a slightly low level
(<82%) for a period of 10 minutes.
Service soon
The concentrator requires servicing
at the earliest convenience. The
concentrator is operating to specifi-
cation and may continue to be used.
Page 30 of 356
What To Do
Contact your equipment provider
to arrange for service and/or order
new columns from the manufacturer.
Wait a few minutes to see if the oxy-
gen concetration improves (alarm
will clear). If condition persists, a
secondary alarm will sound. Follow
the instructions for that alarm or
contact your equipment provider. If
alarm occurs frequently at start up,
this may indicate that maintenance
(column replacement) will soon be
required.
What To Do
Attach an external power supply
turn off and insert a fully charged
battery.
If condition persists, contact your
equipment provider.
Contact your equipment provider to
arrange for service.
96-12100-01-01-A