no battery or the device is not connected to AC or DC power, the alarms will not activate because there is no power.
With the battery connected, a power loss lasting less than 30 seconds will have no effect on the alarm system.
IMPORTANT: If multiple alarm conditions are detected, the highest priority alarm will be displayed.
IMPORTANT: Failure to respond to the cause of an alarm condition for low, medium and high priority alarms poten-
tially will result in discomfort or reversible minor injury only and develop within a period of time sufficient to switch
to a backup source of oxygen.
8.5.1 ALARM LOG
The device maintains a patient accessible alarm log that allows for the last alarm to be accessed and viewed on
the LCD (except for the no-breath-detect, check cannula, battery low / attach plug and batter empty / attach plug
alarms). The alarm log is retained in memory after the device experiences a total loss of power. To access the alarm
log, ensure the concentrator is plugged and turned off. Then hold the plus (+) button for 5 seconds. Alternatively,
the alarm log can be found in the Advanced Tab of the Inogen Connect App under Error Recall.
Once a new alarm is activated the new alarm overwrites the previous alarm. The alarm log is retained in memory
after the device is powered down. The time elapsed since the error occurred is displayed with the last alarm on the
alarm log. The device also maintains a service and repair alarm log that is not accessible by the patient.
8.5.2 INFORMATIONAL SIGNALS (LEVEL 1)
The following notification icons are accompanied by a single, short beep.
Display Icon
©2022 Inogen, Inc. All rights reserved.
Description
Power supply failure or loss of
external power
The battery has stopped charging
and the device has switched to
battery power. Eventually the battery
will be depleted.
Battery hot
Remove battery to cool.
Battery error
Check battery.
Page 29 of 356
What To Do
Plug in the power supply to continue
charging the battery.
The battery needs to be removed
and must be cooled before reuse.
Check the connection of your battery
and ensure that it is properly attached
and latched to the concentrator. If the
battery error persists with same bat-
tery, stop using the battery and switch
to a new battery or remove the battery
and operate the concentrator using an
external power supply.