Inogen Rove 6 User Manual page 47

Portable oxygen concentrator system
Hide thumbs Also See for Rove 6:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
15.3.1 GUIDANCE AND MANUFACTURER'S DECLARATION –
ELECTROMAGNETIC IMMUNITY:
The Concentrator is intended for use in the electromagnetic environment of home, institution, vehicle, train, air-
plane, boats and other transport modalities. The user of the concentrator should make sure it is used in such
an environment. During the immunity testing specified below the Rove 6 will continue to deliver oxygen within
specification.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6Vrms ISM and
amateur frequencies
Radiated RF
10V/m
IEC 61000-4-3
80 MHz
to 2.7 GHz
Electrostatic
± 8 kV contact
discharge (ESD)
± 2, 4, 6, 8 and
IEC 61000-4-2
15 kV air
Electrical fast
± 2 kV for power
transient/burst
supply lines
EC 61000-4-4
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
Voltage dips, short
0% UT for 0.5 cycle at
interruptions and
0 °, 45 °, 90 °, 135 °,
voltage variations
180 °, 225 °, 270 °, and
on power supply
315 °.
input lines
0% UT for 1 cycle
IEC 61000-4-11
70% UT for 25/30 cycle
0% UT for 200/300 cycle
Power frequency
30 A/m
(50/60 Hz)
Magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. main voltage prior to application of the test level.
©2022 Inogen, Inc. All rights reserved.
Electromagnetic Environment Guidance
The Rove 6 Portable Oxygen Concentrator is suitable for the
electromagnetic environment of typical home, institution,
vehicle, train, airplane, boat and other transportation
environments.
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Mains power quality should be that of a typical home,
institution, vehicle and other transportation and mobile
environments. If the user of the Rove 6 requires continued
operation during power mains interruptions, it is recommended
that the device be powered from an uninterrupted power
supply.
Power frequency magnetic fields should be at levels
characteristic of a typical home, institution, vehicle and various
mobile environments. Power frequency magnetic fields from
common appliances in the home are not expected to affect
the device.
Page 47 of 356
96-12100-01-01-A

Advertisement

Table of Contents
loading

This manual is also suitable for:

Is-501Io-501

Table of Contents