arrows mean a larger (or coarser) movement of the waveform in that direction. For
Window Time Shift, align the start of the waveform with the gray vertical line. For First
Break, line up the start of the waveform with the red vertical line.
4. Press Stop
5. Repeat Steps 1-3 for Ports 3&4
6. Press Save
7. If there is more than one NIC-500, select the next NIC-500 from the drop-down menu at
the bottom of the screen and repeat Steps 1-6.
For the NIC500-X (Figure 4-31):
The process to Scope the NIC500-X will depend on the system configuration (number of
sequences and number of firing groups).
1. Select the first firing group and the first channel.
2. Press Auto. The system will initiate a search to attempt to set the Window Time Shift for
the selected firing group and the First Break for the selected channel. The Window Time
Shift will align the start of the waveform with the gray line. The First Break will line up the
start of the waveform with the red vertical line.
Figure 4-31: Scope mode for NIC-500X
48
SPIDAR Software