Service mode

Network settings

Enable network
1.
Select Advanced service in the Service mode.
2.
Select Software upgrade and Network settings.
3.
Supply appropriate values for the network settings into the fields and apply.
(The values shall be supplied by the network administrator at the hospital.)
IP address
Netmask
Default router
4.
Select y to enable the network. The workstation will be turned off.
5.
Connect the network cable.
6.
Switch on the workstation.
Pingtest
A pingtest checks that the default router is alive, it can be performed when the network is
enabled.
1.
Select Pingtest in the Network settings dialog.
NOTICE
To test if a host is alive or not, write the IP address temporarily in the field for Default
router and perform a pingtest.
Siemens AG
Medical Solutions
Advanced Service
NETWORK SETTINGS
Network enabled for this host, opd1190:
IP address:
Netmask:
Default router:
Action: Close
Close/Help/Apply/Pingtest
Fig. 17 Network settings
Address for the system in the network, written as four decimal
numbers separated by periods, e.g. 123.123.123.11.
No initial 0, e.g. 10.10.10.10 is OK but 010.010.010.01 is
not OK.
Netmask used in the network if IP standard subnetting is used.
The IP address of the default router, if this is used in the network.
No initial 0, e.g. 10.10.10.10 is OK but 010.010.010.01 is
not OK.
Some alive routers do not reply to pingtest, check with network
administrator.