9 - 4
Camera
Camera not responding
•
No contact during start-up. Retry.
•
Check that MAMMOMAT and Biopsy controller are powered on.
•
Power off and reboot Opdimar workstation.
•
View error log, if message cannot reach CCD camera on/dev/CCD0 appears, check that
D101 is properly connected inside the workstation.
•
Check cables.
•
Check biopsy controller, fuses and power.
•
Change D100/D101.
•
Change D200.
Cannot use camera
•
Hardware failure in CCD camera. Retry.
•
Check that MAMMOMAT and Biopsy controller are powered on.
•
Power off and reboot Opdimar workstation.
•
Check cables.
•
Check that board D100/D101 is properly connected.
•
Change D100/D101.
•
Change camera.
Opdimar
Service
Register 5
SPB7-230.061.11
Rev. 02
Troubleshooting guide
Page 4 of 14
07.03
CS PS 24
9
9
9
Siemens AG
Medical Solutions