Replacement of fuse in mains filter
The mains filter (3/Fig. 6) contains a fuse. To change fuse, proceed as follows:
1.
Remove the fuse holder (1/Fig. 6) from the filter using a screwdriver and remove
the old fuse (2/Fig. 6).
2.
Install a new fuse in the fuse holder.
3.
Insert the fuse holder into the filter according to Fig. 6.
CAUTION
Fig. 6 Fuse in mains filter
Replacement of fuses on printed circuit board, D200
Fuses on the board D200 in the biopsy controller can be exchanged. The location of the
fuses is shown in MAMMOMAT 1000/3000 Nova - Opdimar Wiring Diagram.
Replacement of EPROMs
To replace the EPROMs of the board D200 in the biopsy unit, proceed as follows:
1.
Use PROM extractor to remove the EPROM.
2.
Install the new EPROM.
Opdimar
Service
Removal and replacement of sub-assemblies
The new fuse must be placed in the same position as the old one
and the marks on the fuse holder and the mains filter must coin-
cide, see Fig. 6.
220-240V
Register 5
SPB7-230.061.11
Rev. 02
3
1
Page 6 of 14
07.03
CS PS 24
2
Siemens AG
Medical Solutions