9 - 10

Network problem

DICOM (option), checklist:
A RJ-45 network connector is needed.
If the distance between the Opdimar workstation and the network connector is longer
than 3,5 m (3,5 m cable included in the package) a UTP-5 cable of maximum 50 m is
needed.
The license key to enable the DICOM (option) is written on the invoice and the dispatch
note (if not, please contact Siemens-Elema AB mammography department in Solna,
Sweden for further information).
To be able to receive the optional DICOM license key the mammography logistics
department will request for the serial No. and the host ID for the workstation in question.
How to receive these parameters is described in Instruction for obtaining the systems
host ID and hostname on Page 5 - 15.
The Opdimar system needs the following information to be connected to the network:
IP address: (e.g. 123.123.123.11)
Netmask: (e.g. 255.255.255.0)
Default router: (e.g. 123.123.123.22)
NOTICE
For further information see DICOM Conformance Statement for Opdimar ASW 2.0 which
is available on Intranet.
Network information about the DICOM nodes, which will be connected to the Opdimar
system.
The following information for all DICOM nodes is necessary:
Hostname: (e.g. opdxxxx)
IP address: (e.g. 123.123.123.11)
DICOM Port: (e.g. 1984 for Opdima)
DICOM AET: (e.g. opdxxxx_SRV)
NOTICE
xxxx is the serial No. for the workstation. The serial No. is labelled on the cover of the
main unit (last four digits). The last four digits of the hostname normally consists of the last
four digits of the serial No. This will not be the case if there has been a hostname change
due to a workstation exchange.
General Information about networking
Project Manager Handbook:
Opdimar
Service
The IP address and the default router may not contain any initial
0, e.g. 10.10.10.10 is OK but 010.010.010.01 is not OK.
For Opdimar, hostname, DICOM Port and DICOM AET are fixed.
Register 5
SPB7-230.061.11
Rev. 02
Troubleshooting guide
CS home page, http://cs.med.siemens.de
For Service
Planning
Project Manager Handbook
Page 10 of 14
07.03
CS PS 24
9
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mammomat nova opdima 3000

Table of Contents