Err700 (Windows File Sharing Error); Err703 (Ic Error); Err704 (Dhcp Error) - Nidek Medical NT-510 Service Manual

Non contact tonometer
Hide thumbs Also See for NT-510:
Table of Contents

Advertisement

XNT5P*RDA002E

5.2.22 ERR700 (Windows file sharing error)

Is the LAN cable connected to the LAN connector?
Yes
Is the network setting correct?
Yes
Contact the network administrator.

5.2.23 ERR703 (IC error)

Turn off and on the power switch of the device.
Is the problem resolved?
No
Replace the main board (18536-BA01)
[p92]).

5.2.24 ERR704 (DHCP error)

The IP address cannot be obtained.
As the DHPC server may not operate normally, contact the network administrator.
Or, set the DHCP connection to OFF and set the IP address and subnet mask
8.11
[p202]).
No
Connect the LAN cable.
No
Correct the network setting
Yes
(see 7.2.1
34
(see 8.11
[p202]).
End