7.

Troubleshooting

Instrument behavior
• Water target temperature
cannot be reached after long
time heating; the red light
blinks all the time.
• Red light is always on; the
heating stops.
• Device cannot be switched on.
• Water leakage on table.
HistoCore Water Bath
User action
1. Clean the water tray completely, especially the internal and
external surface of the bottom part.
2. Switch off and then switch on the device.
3. If the problem cannot be solved by the first 2 steps, contact
customer service.
1. If water temperature is higher than the target temperature, wait
till water cools down.
2. Switch off the instrument and wait for 30 minutes, then switch
on the device.
3. If the problem cannot be solved by the first 2 steps, contact
customer service.
1. Check the fuses integrity and replace both of them.
2. If the problem cannot be solved by fuses replacement, contact
customer service.
Check the water tray to see if there's water leakage and contact
customer service.
Troubleshooting
7
25