Troubleshooting

Stand-alone operation
If the Ivesta 3 (Integrated camera) is configu-
red for 4K resolution and attached to a Full-
HD monitor with only 1,080 p resolution, the
microscope should detect this and switch auto-
matically to the necessary resolution.
However, if no image appears and the screen
reports an error message like "out of range"
or something similar, you can reset the micro-
scope resolution to 1,080 p.
Ivesta 3 Series User Manual
How to reset the
Ivesta 3 (Integrated camera)?
1. Power the microscope off.
2. Unplug the USB mouse.
3. Restart the microscope.
The Ivesta 3 (Integrated camera) will
restart in Full-HD 1,080 p mode and a live
image will appear.
4. Plug in the USB mouse again and configure
the correct live image resolution.